I repaired a Fluke 45 benchtop multimeter I found in a surplus store.
Fortunately the service manual is extremely comprehensive and it turned out to be R5, (a 3.5Kohm, 5W, 1% resistor).
According to the service manual is not a component which affects calibration so I was able to replace it and I now have a wonderful benchtop meter.